Position the 4 numbered buckets in their corresponding positions on the rotor. Each
bucket should pivot freely in its slot.
Place the inserts corresponding to the tubes to be centrifuged in the four buckets.
The buckets must be symmetrically distributed.
For better balance, each numbered bucket must be placed in the corresponding position
marked on the rotor.
Each position on the rotor must be equipped with its bucket.
Do not use a rotor without its complete set of four buckets.
Maximum loads according to the type of rotor are specified in the annex.
The centrifuge tolerates an imbalance of 8 grams.
If the imbalance is higher than the tolerated threshold, an imbalance detection system
stops the centrifuge automatically.
Key logo and error message "02" appear. The rotor stops in free wheel mode without
braking system.
Excessive imbalance is likely to damage the turning components and/or the centrifuge.
Follow loading instructions referring to corresponding rotor description in § 9 Annex 1:
Accessory Specifications.