Original Instructions Rev.1.0 User Manual – Isa 23/ 42
Care and Maintenance
All maintenance actions on this equipment must be performed by a qualified
technician who is familiar with the proper maintenance procedures required for
this equipment.
The device and its accessories may be contaminated.
Apply decontamination measures required before any maintenance.
Turn off the centrifuge power switch. Ensure you lock the power switch when
working on the unit, to avoid any accident from a third party.
Special care must be provided to maintain the original terms of safety and
performance, to extend the life of accessories. A visual inspection can identify any
signs of wear. An annual inspection may be requested according to relevant
regulations
Care and Cleaning
Users should not use cleaning or decontamination methods different from those recommended by the manufacturer, as
they may cause damage to the unit.
The following products and materials are allowed for maintenance:
Distilled water,
70% IPA, peroxides or quaternary ammonium surface disinfectants.
Lint-free cloth
Soft non-metallic brush
Lanolin or silicone spray
Forbidden products and materials:
Chlorine-containing products (Bleach, chlorides, ...)
Saltwater
Wire brush
Perform regular cleaning of the following parts to prevent any risk of corrosion in case of persistent impurity, and to ensure
the hygiene of the working equipment.
External parts lid & keyboard:
Use fluids and tissues impregnated with hydro-alcoholic solution or quaternary ammonium
Check that the air vents are not obstructed.
Centrifuge chamber:
Remove the rotor before cleaning the tank.
Dry thoroughly after cleaning.
Do not use fluffy fabric.
Replace the rotor after cleaning the tank.