Originale Instructions- Rev.1.3 User Maunal– Loreena 14/ 68
2.3. Electrical environment
The centrifuge requires must be 230V/50Hz or 115V/60Hz, single phase, (current rating varies per centrifuge model,
reference Electrical/Environmental Requirements).according to the manufacturer sheet on the back of the unit.
The centrifuge must be plugged into an outlet with protective earthing connection with the standard power cord.
The electrical outlet into which the centrifuge is connected should be readily accessible for maintenance purposes.
Do not position the centrifuge to prevent access to the power cord.
The power cord plug serves as the main disconnect device and should remain readily accessible. The electrical outlet should
be on its own branch circuit, protected with a circuit breaker at the distribution panel near the centrifuge.
2.4. Unpacking
Remove the cardboard packaging made of 2 parts (a lid and a bottom).
For floor model centrifuge, let a minimum of 1 meter space behind the wooden ramp lied on the floor before getting down
the device.
The LOREENA Centrifuge was packed with great care to prevent transport hazards. Before disposing of the packaging of
the centrifuge, and its spinning accessories, please make sure you are in possession of the following:
2.5. Packing list
Manual release key (shape
according centrifuge model)
Centrifuge models :
1) Ventilated or
2) Refrigerating or
3) Floor refrigerating