Originale Instructions- Rev.1.3 User Maunal Loreena 40/ 68
5.4. Checking and testing accessories
Check that no trace of corrosion appears. Check also the absence of any
mechanical damage (traces of impact, corrosion, cracking), or wear on the
rotor and bucket.
In order to ensure optimal performances, once the accessories have reached half of their theoretical life time, we recommend
inspecting them. This life time is different from one accessory to the other, and the value is engraved on the accessory (i.e.:
Max cycles: 55,000 means that the accessory has a theoretical life time of 55,000 cycles)
For safety reasons, the use of accessories that have reached the end of their life cycle is prohibited. These components must
be replaced.
5.5. Storage of accessories
Do not store the accessories on a wire rack. Oxidation build-up may appear and show premature wear of accessories
5.6. Annual maintenance
A periodic check of the centrifuge and accessories is required. Checks must also be carried out in accordance with the
regulations in force: Functional (lubrication of the drive shaft, the lock, readability of labels), security (continuity of the
protective ground connection, checking & maintenance of rotors and accessories) and centrifuge performance checks are
recommended at least once a year.
Replacement of consumables (grease, provided)
Check that the grounding of the device is always properly secured by its power cable and the wall outlet it is connected to.
The operator is not authorized to access the internal part of the equipment. An AFI Centrifuge authorized technician will
ensure maintenance.