Originale Instructions- Rev.1.3 User Maunal– Loreena 17/ 68
3.2. Levelling (Floor model)
Mandatory operation :
Before using floor model centrifuge, adjust the level setting pads,
following the procedure supplied onto the package box
(Document Reference AP3.001 Unpacking & installation).
Use supplied pads.
3.3. Rotor assembly
Clean and lubricate the motor shaft by depositing a thin film of
the food grade grease supplied.
Place the rotor above the drive shaft.
Slide the rotor on the drive shaft.
A click indicates that the rotor is locked.
Clean and lubricate the pins by depositing a thin film of the food
grease supplied.
3.4. Position the buckets
Swing-out rotors are used with round buckets or micro titration
buckets.
Place the buckets in the corresponding numbered locations.
3.5. First powering
Activate the switch on the bottom right of the control center.
On/Off Switch position
Under Control Center