Originale Instructions- Rev.1.3 User Maunal– Loreena 24/ 68
Centrifugation starts.
The progression bar appears
A green tick appears when the rotor stops.
Acquit pressing onto the screen or start/stop button.
Press lid button to open.
4.7. Interrupt centrifugation
During a centrifugation, press the Start/Stop button.
The rotor will start slowing down until complete stop
A green tick appears when the rotor stops.
Acquit pressing onto the screen or start/stop button.
Press lid button to open.
4.8. Open after an electric power failure / Emergency Stop
This operation must be performed by a person trained in centrifugation-
related risks.
In case of power failure, the lid cannot be opened by the engine unlocking
control.
You must wait 15 minutes for the rotor to come to a complete stop before
opening manually.
Floor model :
Turn off the power by pressing the switch to the “0” position.
Insert the wrench provided fully into the round cavity on the left side.
Turn it clockwise to release the lid.
Remove the wrench.
Unload samples.