Originale Instructions- Rev.1.3 User Maunal Loreena 25/ 68
1. Benchtop model :
Insert the flat wrench provided into the vertical slot in the blue
part.
2. The wrench must drive an axis inside the centrifuge.
3. Raise up the wrench to start the lid releasing.
4. Remove and upturn the wrench.
5. Insert back the wrench and raise up for complete lid releasing.
Following a manual release, the wrench must be removed from its housing to prevent accidents related to the rotation
of the wrench during the activation of the locking mechanism. In addition, the security indicator label must be
replaced during a safety check carried out by an authorized technician.
4.9. Precooling
On the Refrigerating model, a "Pre-cooling" cycle prior to centrifugation allows to cool the centrifuge down to 4 degrees
Celsius so it is ready for temperature sensitive samples.
Visual
Instruction
Control
cooling
Select the pre-cooling program:
Rotor: Standard Precool
This program is set to cool the centrifuge and accessories
down to 4°C during a 22 minute cycle at 2500 RPM.
Acceleration and Deceleration slope are both set on 9 and
cannot be modified.
Air: Smart Precool
This program is set to cool the centrifuge until the probe
temperature reaches 4C and will stop when the temperature
is reached.
*Select Rotor or Air in user preferences
Start the cycle by pressing
The temperature inside the bowl will show
on the display
4.10. Prewarming
On the Refrigerating model, a "Pre-warming" cycle prior to centrifugation allows to warm the centrifuge so it is ready for a
protocol at ambient temperature or warmer (especially after a cycle at a colder temperature).
Visual
Instruction
Control
2
2
1
3