Originale Instructions- Rev.1.3 User Maunal– Loreena 7/ 68
1.3. Prohibitions:
The centrifuge is not designed for the following uses and environments.
The following are prohibited:
Flammable, explosive, toxic and radioactive materials
Materials which may react, causing a hazard
Contaminated materials not contained in an aerosol-barrier container
Rotating rotors and accessories that have exceeded their maximum use.
Rotating rotors and accessories that are not maintained according to the instructions in this manual,
or showing signs of wear and/or corrosion.
Rotating rotors and accessories not supplied by AFI Centrifuge.
Explosive environments
Radioactive environments
Only operate a centrifuge after taking all necessary safety measures.
Rotors and buckets should be removed from use in case of mechanical fault,
trace, or corrosion. These elements have a life cycle duration, engraved on
their visible part: To maintain safe conditions, it is imperative to replace them
when the recommended duration is reached.