For arbitrary waveforms used as the carrier waveform, the maximum
frequency depends on the number of points specified in the waveform.
The five built-in arbitrary waveforms can be output at a maximum
of 5 MHz (be sure to note the restrictions below).
For sine, square, and arbitrary waveforms (does not apply to ramp
and triangle waveforms), the relationship between the carrier
frequency and the minimum burst count is shown below.
If you attempt to set the carrier frequency to a value that is not valid,
the function generator will automatically adjust the frequency to the
maximum value allowed with the present burst count.
From the front panel, “
MAX VALUE is displayed and the carrier
frequency is adjusted. From the remote interface, a -221, “Settings
conflict” error is generated and the carrier frequency is adjusted.
Number of Arb Points
8 to 8,192 (8k)
8,193 to 12,287 (12k)
12,288 to 16,000
Minimum Frequency
10 mHz
10 mHz
10 mHz
Maximum Frequency
5 MHz
2.5 MHz
200 kHz
Carrier
Frequency
10 mHz to 1 MHz
>1 MHz to 2 MHz
>2 MHz to 3 MHz
>3 MHz to 4 MHz
>4 MHz to 5 MHz
Minimum
Burst Count
1
2
3
4
5
For sine, square, and
arbitrary waveforms only.
3
Chapter 3 Features and Functions
Burst Modulation
85

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals