30 Agilent 35670A Supplement
4 Service Guide
What to do after replacing an assembly
The rows referring to A7 CPU, A8 Memory, and A9 NVRAM are
replaced with:
Note that the frequency adjustment is no longer needed.
Disregard the last two rows of the table referring to adjustments
of the Display and the DC- DC Converter (now a Power Inverter).
There are no adjustments for these items.
Page in original guide 6-5
Assembly Replaced Required Adjustment Required Performance Test
A17 CPU No No