Gas Requirements and Configurations
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 49
Table 15. Compressed air specification for the nitrogen generator and accessories
Intended use POP gas Nitrogen
generator
plasma gas
Monochromator
nitrogen
generator purge
Organics Monochromator
air purge
Quality ISO 8573-
1:2010 CLASS
1.4.3
ISO 8573-
1:2010 CLASS
8.4.3
ISO 8573-1:2010
CLASS 8.4.3
ISO 8573-
1:2010
CLASS 1.4.3
ISO 8573-1:2010
CLASS 1.4.3
Required
flow (SLPM)
25 115 115 1.5 10
Required
pressure
(kPa)
450–600 620–750 620–750 450–600 450–600
Max output
flow (SLPM)
n/a 25 10 n/a n/a
Output
pressure
(kPa)
n/a 450–600 450–600 n/a n/a
The following pages detail some the possible 4100 MP-AES and
accessory configurations along with their associated gas
requirements. Use these as an example for calculating/determining
your gas requirements.

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals