Gas Requirements and Configurations
52 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Figure 3. Agilent 4100 MP-AES, N
2
plasma gases supplied from nitrogen generator, argon gas
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with N
2
to allow determination of sulfur.
Total Compressed Air = 115+115+25 = 255 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
MP-AES
MP-AES Plasma
Gas Box
Agilent
Nitrogen
Generator
#1
Oil Free Air
Compressor
and Dryer
Agilent
Nitrogen
Generator
#2
EGCM
Filter/ Regulator