Gas Requirements and Configurations
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 53
Figure 4. Agilent 4100 MP-AES, N
2
plasma gases supplied from nitrogen generator, argon gas
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, Monochromator
Air Purge supplied from air compressor to keep the monochromator at positive pressure and
exclude any dust or vapors.
Total Compressed Air = 115+10+25 = 150 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
Oil Free Air
Compressor
and Dryer
Agilent
Nitrogen
Generator
MP-AES
MP-AES Plasma
Gas Box
Mono
Air
Purge
Filter/ Regulator