Gas Requirements and Configurations
54 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Figure 5. Agilent 4100 MP-AES, N
2
plasma gases supplied from nitrogen generator #1, argon gas
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with N
2
from nitrogen generator #2 and to supply air from the air
compressor for analysis of organics.
Total Compressed Air = 115+115+25+1.5 = 256.5 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
Oil Free Air
Compressor
and Dryer
Agilent
Nitrogen
Generator
#1
MP-AES
MP-AES Plasma
Gas Box
EGCM
Agilent
Nitrogen
Generator
#2
Filter/ Regulator