Gas Requirements and Configurations
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 55
Figure 6. Agilent 4100 MP-AES, N
2
plasma gases supplied from nitrogen generator, argon gas
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with air from the air compressor and to provide air for analysis of
organics.
Total Compressed Air = 115+10+25+1.5 = 151.5 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
EGCM
Agilent
Nitrogen
Generator
Oil Free Air
Compressor
and Dryer
MP-AES Plasma
Gas Box
MP-AES
Filter/ Regulator