Gas Requirements and Configurations
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 59
Figure 10. Agilent 4100 MP-AES, N
2
plasma gases supplied from cylinder/Dewar system, argon
gas supplied from Onboard Argon Supply, POP gas supplied from the air compressor. Air supplied
to the Monochromator Air Purge from the air compressor.
Total Compressed Air = 25+10 = 35 SLPM
Total Nitrogen (99.5%) = 25 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
MP-AES
MP-AES Plasma
Gas Box
Oil Free Air
Compressor
and Dryer
Mono
Air
Purge
Filter/ Regulator
Nitrogen
Supply