Gas Requirements and Configurations
60 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Figure 11. 4100 MP-AES, N
2
plasma gases supplied from cylinder/Dewar system, argon gas
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with nitrogen from a nitrogen generator and provide air from the air
compressor for organics analysis.
Total Compressed Air = 115+25+1.5 = 141.5 SLPM
Total Nitrogen (99.5%) = 25 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
Oil Free Air
Compressor
and Dryer
MP-AES
MP-AES Plasma
Gas Box
Agilent
Nitrogen
Generator
EGCM
Filter/ Regulator
Nitrogen
Supply