Agilent 4155C/4156C VXIplug&play Driver Users Guide, Edition 4 2-11
Driver Function Reference
Parameters
Table 2-8 PGU output voltage, resolution, and compliance by range
Table 2-9 PGU pulse period, pulse width, and delay time
Table 2-10 PGU leading transition time and trailing transition time
Output range
(actually used)
Output voltage
a
a. Parameter name may be base, bias, peak, stress, value, start, stop, and
so on.
Setting
resolution
comp value
b
b. Set any value (e.g. 0.1) though the compliance is always 100 mA.
20 V 0 |V| 20 V 4 mV ±100 mA
40 V 0 |V| 40 V 8 mV ±100 mA
period value width value
delay value
a
a. The value must be 0 delay period value.
Resolution
2.0 μs to 100.0 μs1.0 μs to 99.9 μs 0 to 100.0 μs0.1 μs
100 μs to 1000 μs1 μs to 999 μs 0 to 1000 μs1 μs
1.00 ms to 10.00 ms 0.01 ms to 9.99 ms 0 to 10.00 ms 10 μs
10.0 ms to 100.0 ms 0.1 ms to 99.9 ms 0 to 100.0 ms 100 μs
100 ms to 1000 ms 1 ms to 999 ms 0 to 1000 ms 1 ms
1.00 s to 10.00 s 0.01 s to 9.99 s 0 to 10.00 s 10 ms
Leading transition time
(rise)
Trailing transition time
(fall)
Resolution
100 ns to 1000 ns 100 ns to 1000 ns 1 ns
0.50 μs to 10.00 μs0.50 μs to 10.00 μs10 ns
5.0 μs to 100.0 μs5.0 μs to 100.0 μs100 ns
50 μs to 1000 μs50 μs to 1000 μs1 μs
0.5 ms to 10.00 ms 0.5 ms to 10.00 ms 10 μs

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals