Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4 2-37
Driver Function Reference
hp4156b_setSweepSync
If mode=2, this is the stress time (in seconds). 500 μs to 655.0 s.
period Pulse period (in seconds). 2 μs to 10.0 s. Only for pulse stress.
The period value must match the width and delay values of the
hp4156b_addStressSyncPulse function. See Table 2-9.
For mode=1, if you set the automatic abort function by using the
hp4156b_stopMode function, the pulse output must be more
than 10 seconds (duration × period > 10 s).
NOTE Pulse Count and Pulse Period
Pulse count (duration) and pulse period (period) settings are effective for both
PGU1 and PGU2. If you use a PGU as a stress source set by this function and
another PGU as a pulse source set by the hp4156b_forcePulse function, check the
setting of these parameters. The parameter values must be the same as shown below.
hp4156b_setStress’s duration (pulse count) = hp4156b_forcePulse’s count
hp4156b_setStress’s period = hp4156b_forcePulse’s period
hp4156b_setSweepSync
This function specifies synchronous sweep source and sets the parameters. The
synchronous sweep source will be the additional staircase sweep source for the
staircase sweep measurements, the pulsed sweep measurements, or the staircase
sweep with pulsed bias measurements. The hp4156b_setIv or hp4156b_setPiv
function must be executed before this function.
For the staircase sweep measurements, the output synchronizes with the staircase
sweep output by the hp4156b_setIv function.
For the pulsed sweep measurements, the output synchronizes with the pulsed sweep
output by the hp4156b_setPiv function.
For the staircase sweep with pulsed bias measurements, the output synchronizes the
staircase sweep output by the hp4156b_setIv function and the pulse output by the
hp4156b_setPbias function.
Syntax ViStatus _VI_FUNC hp4156b_setSweepSync(ViSession vi, ViInt32 channel,
ViInt32 mode, ViReal64 range, ViReal64 start, ViReal64 stop, ViReal64 comp,
ViReal64 p_comp);
Parameters vi Instrument handle returned from hp4156b_init( ).

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals