Exercise 1 – Develop an acquisition method
Task 3. Find optimum fragmentor voltage for maximum response
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide 17
3 Set up a qualitative method to
view the EIC data automatically.
• Open the data file
Sulfa_SIM60.d or your own
iiiSulfa_SIM60.d, where iii are
your initials.
• In the Method Editor, add in the
EICs corresponding to the
precursor ion masses of 271,
279, 285, and 311.
• Save the method as iiiExercise1,
where “iii” are your initials.
a Click File > Open Data File.
The system displays the Open Data
File dialog box
b Select either Sulfa_SIM60.d or
iiiSulfa_SIM60.d, and click Open.
c Click Method > Method Editor or
View > Method Editor.
The system displays the Method
Editor window.
• The Qualitative Analysis program
should be open. If not, see
“Double-click the Qualitative
Analysis icon.” on page 11.
• By default, the Method Editor is a
floating window. The window is
docked for these images. See the
online Help for more information on
floating and docking windows.
Steps Detailed Instructions Comments