Note
The
CW frequencies
c
hec
k
ed
pro
vide
a
minim
um
v
erication
of
pulse
p
erformance.
If
pulse
p
erformance
at
a dieren
t
frequency is
imp ortan
t, rep
eat the
test
at
that
CW
frequency
.
3.
On
the
pulse generator,
set:
Pulse
Width:
50 ns
Pulse
Period:
10
us
(100
kHz)
Pulse
Level:
5V
Disable:
LED
off
(enables
pulse
generator)
4.
On
the
oscilloscop
e,
set:
Channel
3:
Ext Probe
10
Display
On
Volts/Division
15
mV
(Readjust
as
required)
Offset 0V
Timebase:
Time/Division
5
ns
Delay
100
ns
Delay
Reference
At
center
Trigger:
Trigger
Slope
Positive
Probe
Atten
20
Trigger
Level 1.2V
Display:
Display
Mode
Persistence
Display Time
1 sec
Screen Single
Note
On the oscilloscop
e, turn the w
aveform math function on. Then
use the w
aveform math maximum function to determine the
pulse en
velope when making
the follo wing measuremen
ts.
2-70 Performance Tests
Pulse Performance
HP 8360
All manuals and user guides at all-guides.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620A and is the answer not in the manual?

Agilent Technologies 83620A Specifications

General IconGeneral
BrandAgilent Technologies
Model83620A
CategoryBlower
LanguageEnglish