15.
Set
the
synthesizer
to eac
hof
the
cen
ter
frequencies
in
T
able
3-7
.
Change
the p
o
w
er
sw
eep
as
appropriate
to
sw
eep
the
full
ALC
range.
P
erform
an
AM
bandwidth
calibration
and
then
rep
eat
step
14
for
each
corresp onding
calibration
constant.
T
able
3-7.
Center
Frequencies and
Calibration Constants
Cen
ter
F
requencies
(GHz)
Calibration
Constan
ts
5.0 #316 Band
1
9.0 #317
Band
2
19.0 #318
Band 3
25.0
1
#319
Band
4
30.0
1
#320
Band
5
35.0
1
#321
Band 6
45.0
1
#322
Band
7
1
Disregard
if
b
ey
ond
y
our
syn
thesizer's
capabilit
y.
Related P
erformance T
ests
Maximum
Leveled
P
o
w
er
AM
Bandwidth
In
Case
of
Difficulty
Refer
to
the
T
r
oublesho
oting
man
ual.
3-90 Adjustments
Modulator Offset and Gain
HP 8360
All manuals and user guides at all-guides.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620A and is the answer not in the manual?

Agilent Technologies 83620A Specifications

General IconGeneral
BrandAgilent Technologies
Model83620A
CategoryBlower
LanguageEnglish