Figure
2-16.
Pulse
Env
elope Displa
y
ed
Ov
er
5
Divisions
7.
Adjust
the
timebase
dela
y
to
position
the falling
edge
of
the
pulse
near
the
cen
ter
of
the
oscilloscop
e
displa
y
.
8. Measure
the
falltime
the
same
w
a
y
as
y
ou
measured
the
risetime
(see
step
6).
Record
this
value
in T
able 2-18
.
9.
Rep
eat
steps
5
through
8
at
eac
h
synthesizer
frequency in
Table
2-18.
10.
Record
eac
h
w
orst
case
risetime
and
falltime
v
alue
from
T
able
2-18
on
the
test
record.
2-72 Performance Tests