Figure
2-27.
AM
Dynamic
Range
T
est
Setup
Equipment
Sp
ectrum
Analyzer
HP
8566B
Digital
Voltmeter
HP
3456A
F
unction Generator
HP 3325A
P
o
w
er
Meter
HP
436A/7A/8A
P
o
w
er
Sensor
HP
8487A
(HP
83640A/50A)
P
o
w
er
Sensor
HP
8485A
(HP
83620A/22A/23A/24A/30A)
2. On the syn
thesizer, turn on RF p eaking and AM bandwidth
calibration.
Press:
4
USER CAL
5
NNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNN
Tracking Menu
NNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNN
NNN
Peak RF Always
(asterisk on)
4
PRIOR
5
NNNNN
NNNNNNNNNNNNNNNNNNNNNNNNN
NNNNN
AM Cal Menu
NNNNN
NNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
AM BW Cal Always
(asterisk on)
3. Zero and calibrate the po
wer meter/sensor.
4. Connect the equipmen
tassho
wn in Figure 2-27 with the
power sensor
connected to
the synthesizer's RF OUTPUT.
2-108 Performance Tests
AM Dynamic Range
HP 8360