SCPI COMMAND SUMMARY
DIAGnostics:INSTrument:PMETer:ADDRess <num>
DIAGnostics:INSTrument:PMETer:ADDRess?
Sets and queries the GPIB address to use for the power meter during
swept signal generator calibration routines. Allowable values are 0
through 31. *RST or p ower on do es not eect this value. Default is
13. It is defaulted only when memory is initialized.
DIAGnostics:INSTrument:PRINter:ADDRess <num>
DIAGnostics:INSTrument:PRINter:ADDRess?
Sets the GPIB address of the printer to use during some of the
calibration pro cedures when the swept signal generator assumes
GPIB control. *RST and p ower on do not eect this command. The
default is 1. The default value is set at memory initialization only.
DIAGnostics:IORW <num>,<num>
Performs a write to the I/O Device number sp ecied in the rst
<
num
>
and sets it to the value in the second
<
num
>
.
DIAGnostics:IORW? <num>
Reads from the specied I/O device number and returns the resp onse
data.
DIAGnostics:OUTPut:FAULt?
Returns a string of 16, 1s and 0s that are equiv
alent to the fault
display.
Bit 0 = PEAK
Bit 1 = TRACK
Bit 2 = RAMP
Bit 3 = SPAN
Bit 4 = V/GHz
Bit 5 = ADC
Bit 6 = EEROM
Bit 7 = PWRON
Bit 8 = CALCO
Bit 9 = PLLZERO
Bit 10 = PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 = CAL MAN
Bit 14 = TMR CNFLCT
Bit 15 = SEARCH
Operating and Programming Reference S-31