Saving and
Recalling an
Instrument State
The save/recall registers store and access a previously set instrument
state.
For example, set the swept signal generator to sweep from 3 GHz to
15 GHz at a
0
10 dB p ower level, with markers 1 and 2 set at 4.5 and
11.2 GHz.
Press
4
START
54
3
54
GHz
5
.
Press
4
STOP
54
1
54
5
54
GHz
5
.
Press
4
POWER LEVEL
54
0
54
1
54
0
54
dB(m)
5
.
Press
4
MARKER
5
.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Marker M1
4
4
54
.
54
5
54
GHz
5
.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Marker M2
4
1
54
1
54
.
54
2
54
GHz
5
.
Tosave this instrument state in register 1, press
4
SAVE
54
1
5
.Toverify
that the swept signal generator has saved this state:
Press
4
PRESET
5
.
Press
4
RECALL
54
1
5
.
Press
4
MARKER
5
.
The active entry area displays:
--> RECALL REGISTER: 1 RECALLED
Notice the sweep end points, p ower level, and the asterisks next to
the marker 1 and 2 key lab els.
You can save instrument states in registers 1 through 8. Register
0saves the last instrument state b efore power is turned o. When
power is turned on, register 0 is automatically recalled.
1-16 Getting Started Basic