SCPI COMMAND SUMMARY
DIAGnostics:RESult?
Returns the following information:
<result>, <test failure>, <manual entry point>
where,
<result>
is one of the following:
0 Diagnosis successful.
1 Cannot diagnose; full selftest must b e executed rst.
2 No failures found|all selftests passed.
0
1 Cannot diagnose; diagnosis routine failed to isolate failure.
(software fault)
<test failure>
is the test number of the most relevant failure
(
0
999 if parameter is not used, as in
<
result
>
of 1).
<manual entry point>
is a string response that identies the
paragraph number in the
Agilent Technologies 8360 B-Series Swept
Signal Generator/8360 L-Series Swept CW Generator Service Guide
to begin the troublesho oting pro cedure.
DIAGnostics:TEST:CONTinue
Causes the selftest execution to continue when paused for raw data
examination. Do es nothing in other conditions.
DIAGnostics:TEST:DATA:DESC?
Returns the description string of the raw data examined during a
selftest. It returns \ " in other conditions.
DIAGnostics:TEST:DATA:MAXimum?
Returns the upper test limit for the ra
w data point examined.
Returns \0" in other conditions.
DIAGnostics:TEST:DATA:MINimum?
Returns the lower test limit for the raw data p oint examined.
Returns \0" in other conditions.
DIAGnostics:TEST:DATA:VALue?
Returns the raw data value for the raw data p oint examined.
Returns a \0" in other conditions.
DIAGnostics:TEST:DISable {<num>}1*?|ALL
Prevents the listed selftests from being selected. If
ALL
is sent then
all of the selftests are disabled. *RST causes
DIAG:TEST:ENAB ALL
to
execute.
DIAGnostics:TEST:ENABle {<num>}1*?|ALL
Enables the listed selftests to execute. If
ALL
is sent then all of the
selftests are enabled. *RST causes
DIAG:TEST:ENAB ALL
to execute.
S-32 Operating and Programming Reference