SCPI COMMAND SUMMARY
DIAGnostics:TEST[:EXECute] <num>
The specied selftest is executed. Normal instrument op eration is
suspended and the instrument state is restored upon exiting the
selftest mo de.
DIAGnostics:TEST:LOG:SOURce ALL|FAIL
DIAGnostics:TEST:LOG:SOURce?
Sets and queries the source for the raw data logging.
ALL
species
that all raw data p oints are displayed.
FAIL
selects only those data
points out of the test limits. Both commands are executable in
selftest mo de. After *RST, the setting is
FAIL
.
DIAGnostics:TEST:LOG[:STATe]?
DIAGnostics:TEST:LOG[:STATe] ON|OFF|1|0
Selects and queries the raw data logging ON/OFF switch. Both
commands are executable in selftest mo de.
After *RST, the setting is 0.
DIAGnostics:TEST:LOOP ON|OFF|1|0
DIAGnostics:TEST:LOOP?
Selects and queries the test lo oping ON/OFF switc
h. Both
commands are executable in selftest mo de.
After *RST, the setting is 0.
DIAGnostics:TEST:NAME? [<num>]
Queries the name of a selftest bynumber. If the number is not
specied then an array of all the selftest names is returned.
DIAGnostics:TEST:POINts?
Returns the number of p oints of selftest that is output using
DIAGnostics:TEST:NAME?
or
DIAGnostics:TEST:RESult?
.
DIAGnostics:TEST:RESult? [<num>]
Queries the result of a selftest, bynumber. The resp onse is a string
containing either
Passed
,
Failed
,or
NotRun
.If
<
num
>
is missing,
an array of selftest results are returned.
DIAGnostics:TINT? <num>
A test feature that returns the value passed to it. This is used to test
the GPIB interface.
DISPlay[:STATe] ON|OFF|1|0
DISPlay[:STATe]?
Sets and queries the display ON/OFF switch.
After *RST, the value is 1.