2-36
Making Mixer Measurements
Phase or Group Delay Measurements
Figure 2-24 Connections for a Group Delay Measurement
6. To select the converter type and a high-side LO measurement configuration, press:
7. To view the measurement results on the analyzer’s display, press:
8. To select the format type, press:
9. To make a response error-correction, press:
System
INSTRUMENT MODE
FREQ OFFS MENU
DOWN CONVERTER
RF<LO
FREQ OFFS ON
System
INSTRUMENT MODE
FREQ OFFS MENU
VIEW MEASURE
Format
DELAY
Meas
INPUT PORTS
B/R
Cal
CALIBRATE MENU
RESPONSE
THRU