Menu Reference
Front Panel Menu Reference "System/Inputs" Menu
Agilent E4419B User’s Guide 3-33
■
This softkey is only present on power meters fitted with the 
battery option 001. It is greyed out when the battery is not 
installed.
Press this softkey to produce a screen which shows:
- How much charge the battery is holding
- The approximate running time left under battery power.
- Either, “Using ac power” (if the battery is fully charged), 
“Charging battery”, or “Using battery power” as 
applicable.
This softkey also accesses the Battery menu which allows you 
to set the display backlight to On, Off, or Timed mode. In 
Timed mode, the backlight turns off 10 minutes after the last 
key press. To turn the backlight on again, press any key. While 
operating with ac power connected these keys are grayed out 
and the backlight is permanently on. On instrument preset 
the Backlight is set to “On”
◆
Press this softkey to turn the display backlight permanently 
On.
◆
Press this softkey to set the display backlight to be Off while 
the power meter is operating under battery power.
◆
Press this softkey to select the Backlight Timed mode. When 
the power meter is in this mode and is operating under 
battery power, the display backlight turns off 10 minutes 
after the last key press. To turn the backlight back on, press 
any key.
◆
Pressing this softkey returns the display to the previous 
screen.
Battery
Backlight On
Backlight Off
Backlight Timed
HP4402.book  Page 33  Thursday, November 30, 2000  1:39 PM