48
3 Installing the Autosampler
Flow Connections
Flow Connections
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
valve.
3 Connect the corrugated waste tube to the solvent waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
Parts required # Description
1 Parts from the Accessory kit
Preparations Sampler is installed in the LC system
WARNING
Toxic and hazardous solvents
The handling of solvents and reagents can hold health risks.
When opening capillary or tube fittings solvents may leak out.
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
NOTE
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
complete seat capillary introducing air into the system.

Table of Contents

Other manuals for Agilent Technologies G1329A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1329A and is the answer not in the manual?

Agilent Technologies G1329A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1329A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals