Optimizing Performance
5
Fast Injection Cycle and Low Delay Volume
General Recommendations for Fast Injection Cycle Times
As described in this section, the first step to provide short cycle times are
optimizing the chromatographic conditions. If this is done the autosampler
parameter should be set to:
Overlapped injection mode
Increase of draw and eject speed for large injection volumes
Add at last run a blank, if overlapped injection is used
To reduce the injection time, the detector balance has to be set to OFF.
NOTE
For the last injection of the sequence with overlapped injections it has to be considered
that for this run the injection valve is not switched as for the previous runs and
consequently the injector delay volume is not bypassed. This means the retention times are
prolonged for the last run. Especially at low flow rates this can lead to retention time
changes which are too big for the actual calibration table. To overcome this it is
recommended to add an additional “blank” injection as last injection to the sequence.

Table of Contents

Other manuals for Agilent Technologies G1329A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1329A and is the answer not in the manual?

Agilent Technologies G1329A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1329A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals