7 Maintenance
Simple Repairs
5
Reconnect the sample-loop fitting to the needle fitting. 6 Use Needle Up to lift the needle to a position
approximately 2mm above the seat.
7 Ensure the needle is aligned with the seat. Next Steps:
8 On completion of this procedure: Install the front cover.
9 Select “End” in the maintenance function “Change
Needle”.