Agilent InfinityLab LC Series Vialsamplers User Manual 264
8Maintenance
Exchange the Wash Port Assembly
3 Verify that the needle in the uppermost position before
continuing.
4 Loosen the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
5 Remove the safety cover.
WARNING
Sharp needle
Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
6 Move the wash port into the service position by slightly
lifting it up and then flapping to the left.