2-4 Service Guide N5242-90001
General Product Information PNA Series Microwave Network Analyzers
Analyzer Options Available N5242A
Option 020, Add IF Inputs for Antenna and Millimeter-Wave
This option can be added to any other option combination.
This option enables external IF inputs on the rear panel of the analyzer for applications
such as antenna and millimeter-wave test.
Option 021, Add Pulse Modulator to 1st Internal Source
This option can be added to any other option combination.
This option enables the internal pulse modulator on source 1. An external pulse generator
can be connected through the rear-panel PULSE I/O connector or one of the internal pulse
generators provided with Option 025 can be used. The N1966A pulse I/O adapter is
recommended if using an external pulse generator. See “Analyzer Accessories Available” on
page 2-9.
Option 022, Add Pulse Modulator to 2nd Internal Source
This option can be added to any analyzer with Option 224 or Option 400.
This option enables the internal pulse modulator on source 2. An external pulse generator
can be connected through the rear-panel PULSE I/O connector or one of the internal pulse
generators provided with Option 025 can be used. The N1966A pulse I/O adapter is
recommended if using an external pulse generator. See “Analyzer Accessories Available” on
page 2-9.
Option 025, Add Four Internal Pulse Generators
This option can be added to any other option combination.
This option enables four internal pulse generators. These pulse generators can be used to
control the internal pulse modulators (Options 021 and 022) and internal receiver IF gates.
These internal pulse generators are also available through the rear-panel PULSE I/O
connector to control external pulse modulators. The N1966A pulse I/O adapter is
recommended if using an external pulse generator. See “Analyzer Accessories Available” on
page 2-9.
Option 029, Noise Figure Measurement Capability
This option can be added to Options 219, 224, 419, or 423.
This option provides the additional hardware and firmware necessary to add
high-accuracy noise figure measurement capability to your 2-port or 4-port analyzer.
Option 080, Frequency Offset Measurements
This option can be added to any other option combination.
This option enables the additional firmware necessary to make frequency offset
measurements by enabling the analyzer to set the source frequency independently from
where the receivers are tuned. This ability is important for measuring amplifiers, mixers,
and frequency converters.