EasyManua.ls Logo

Agilent Technologies N5242A - 10 Mhz Frequency Reference Adjustment

Agilent Technologies N5242A
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-36 Service Guide N5242-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
Adjustments N5242A
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network
analyzers 10 MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the 10 MHz Frequency Reference Adjustment
Procedure
NOTE
This adjustment typically adjusts to within ±0.01 ppm.
1. Connect the equipment as shown in Figure 3-13. Connect a GPIB cable between the
network analyzer and the frequency counter.
Figure 3-13 Equipment Setup for the 10 MHz Frequency Reference Adjustment
2. Press UTILITY , then , then , then click
10 MHz Freq Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
Equipment Type Model or Part Number Alternate Model or Part Number
Cable, BNC, 50Ω, 24 inch 8120-1839 Any
Frequency counter 53151A, Option 001 Any that will measure a signal at 10 MHz.
System
Service Adjustments

Table of Contents

Related product manuals