EasyManua.ls Logo

Agilent Technologies N5242A - Removing and Replacing the A18 GPIB Board

Agilent Technologies N5242A
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-24 Service Guide N5242-90001
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing the A18 GPIB Board N5242A
Removing and Replacing the A18 GPIB Board
Tools Required
T-10 TORX driver (set to 9 in-lb)
T-20 TORX driver (set to 21 in-lb)
ESD grounding wrist strap
Removal Procedure
Refer to Figure 7-11 for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to “Removing the Covers” on page 7-6.
3. Remove the A16 power supply assembly. Refer to “Removing and Replacing the A16
Power Supply Assembly” on page 7-20.
4. Remove the A17 CPU board assembly. Refer to “Removing and Replacing the A17 CPU
Board Assembly” on page 7-22.
5. Remove the rear panel. Refer to “Removing and Replacing the Rear Panel” on page
7-56.
6. Slide the A18 GPIB board out the rear of the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to Table 7-2 on page 7-64.

Table of Contents

Related product manuals