4-16 Service Guide N5242-90001
Troubleshooting PNA Series Microwave Network Analyzers
Front Panel Troubleshooting N5242A
Checking the A2 USB Board
To verify proper operation of the USB board:
Connect a known good USB device, such as a USB mouse, to a front panel USB port.
Wait 15 seconds for the analyzer to verify the device connection, and then check the
operation of the USB device.
If the device performs correctly, the USB board is functioning properly.
If the device does not perform correctly, the USB board is faulty. Refer to “Removing the
A2 USB Board” on page 7-10.
A1 Front Panel Interface Board
This assembly performs the following functions:
It routes USB signals between the front-panel USB connector and the A17 CPU board.
The speaker produces the audio output from signals supplied by the A17 CPU board.
It routes key pad commands from the keypad to the A17 CPU board.
It routes display signals from the A17 CPU board to the A3 display assembly.
Checking the Speaker
If no audio is heard:
Verify that the volume is set correctly and the proper sound driver is loaded; do the
following:
Press UTILITY , then , then .
Click on the
Sounds and Audio Devices entry. Follow the normal Windows procedure
to check the sound drivers and volume. If the audio is still not heard, suspect a faulty
speaker. Refer to “Removing the A1 Front Panel Interface Board and Keypad
Assembly” on page 7-10.
Checking the Operation of the Key Pad Commands
To verify the key pad functionality, refer to “Front Panel Keypad and RPG Test” on
page 4-13.
Checking the Display
To verify the display functionality, refer to “A3 Display Test” on page 4-15.