3-8 Service Guide N5242-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
ANSI/NCSL Z540–1–1994 Verification N5242A
ANSI/NCSL Z540–1–1994 Verification
To meet the criteria for ANSI/NCSL Z540-1-1994, perform the preliminary checks and
either system verification or performance tests without stopping to repair or adjust
1
. Refer
to Figure 3-1 for test flow. Print data at the completion of all the tests, even if you are
aware that the analyzer did not pass. If there is a failure, complete the verification before
you troubleshoot, repair, and adjust. After the failure has been corrected, repeat the entire
system verification or performance tests and generate a new set of data.
Figure 3-1 ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage