Service Guide N5242-90001 7-13
PNA Series Microwave Network Analyzers Repair and Replacement Procedures
N5242A Removing and Replacing the A1A3 and Other Front Panel Subassemblies
4. The display cable (item
) can be removed by removing the three screws (item
) that
attach it to the LCD display hold down bracket (item
).
5. The LCD display can be removed by disconnecting the LCD cable from the inverter
board and then removing the four screws (item
) that attach it to the LCD display
hold down bracket (item
). Note the location of the four spacers (item
) before
separating the LCD display from the hold down bracket.
Figure 7-5 A3 Display Assembly and Touch Screen Removal-2
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to Table 7-2 on page 7-64.