Service Guide N5242-90001 7-55
PNA Series Microwave Network Analyzers Repair and Replacement Procedures
N5242A Removing and Replacing the A54 Option 029 Test Port 2 Bridge and the A55 Option 029 Noise Downconverter
Replacement Procedure
1. Reverse the order of the removal procedure.
When reassembling the A54/A55 assembly, lay the assemblies on a flat surface while
installing and tightening the interconnecting cable (adapter) to ensure that these
assemblies will be flush with the side frame when installed.
Torque all cable connections to 10 in-lbs.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to Table 7-2 on page 7-64.