3-18 Service Guide N5242-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
Preliminary Checks N5242A
12. For the Port 2, 3, and 4 Check, connect the cable (with the load attached) to the
respective port and repeat steps 2 through 11.
Figure 3-7 Typical Cable Connector Repeatability Response
If the Cable Connector Repeatability Check Fails
1. Clean the cable and devices, and torque to specification. Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.