Language Dictionary 6
Series N6700 User’s Guide 105
SYSTem:PASSword:FPANel:RESet
This command resets the front panel lockout password to the factory-
shipped setting, which is zero (0). This command does not reset the
calibration password.
NOTE
The front panel password can also be reset to 0 by setting an internal switch on
the unit. This switch will also reset the calibration password to 0. Refer to
Appendix B under “Calibration Switches” for more information.
SYSTem:REBoot
This command causes the instrument to reboot to its power-on state.
SYSTem:VERSion?
This query returns the SCPI version number to which the instrument
complies. The returned value is of the form YYYY.V, where YYYY
represents the year and V is the revision number for that year.
*IDN?
This query requests the power system to identify itself. It returns a
string of four fields separated by commas.
Field Information
Agilent Technologies
N67xxA
0
<A>.xx.xx
Manufacturer
Mainframe model number followed by a letter suffix
Zero or mainframe serial number if available
Revision levels of firmware
*OPT?
This query requests the mainframe to identify any installed options.
A 0 indicates no options are installed.
*RCL <state>
This command restores the power system to a state that was
previously stored in memory locations 0 through 1 with the *SAV
command. All instrument states are recalled except for the following:
The trigger system is set to the Idle state by an implied ABORt
command (this cancels any uncompleted trigger actions).
Calibration is disabled by setting CALibration:STATe to OFF.
All list settings are set to their *RST values.
All status registers are set to their PRESet values.
NOTE
The device state stored in location 0 is automatically recalled at power turn-on
when the Output Power-On state is set to RCL 0.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals