Language Dictionary 6
Series N6700 User’s Guide 107
Trigger Subsystem
The Trigger subsystem consists of the Abort, Initiate, and Trigger
commands.
Abort commands cancel any triggered actions.
Initiate commands initialize the trigger system. This enables the
trigger system to receive triggers.
Trigger commands control the remote triggering of the power system.
They specify the trigger source for the transient and the
measurement system and also generate software triggers.
ABORt:ACQuire (@<chanlist>)
ABORt:TRANsient (@<chanlist>)
These commands cancel any measurement or transient trigger
actions presently in process. The two WTG bits in the Status
Operation Condition register are also reset. These commands are
executed at power-on and upon execution of *RST.
INITiate[:IMMediate]:ACQuire (@<chanlist>)
INITiate[:IMMediate]:TRANsient (@<chanlist>)
These commands control the enabling of both measurement and
transient triggers. When a trigger is initiated, an event on a selected
trigger source causes the specified triggering action to occur. If the
trigger system is not initiated, all triggers are ignored.
It takes a few milliseconds for the power supply to be ready to
receive a trigger signal after receiving the INITiate command, and it
can take longer if the value of SENSe:SWEep:OFFSet is negative. If a
trigger occurs before the trigger system is ready for it, the trigger will
be ignored. The user can test the WTG_MEAS bit in the operation
status register to know when the power supply is ready to receive a
trigger after initiating.
INITiate:CONTinuous:TRANsient ON|OFF, (@<chanlist>)
INITiate:CONTinuous:TRANsient? (@<chanlist>)
This command continuously initiates the output trigger system. The
enabled state is ON (1); the disabled state is OFF (0). With continuous
triggering disabled, the output trigger system must be initiated for
each trigger using the INITiate:TRANsient command.
The *RST value = OFF.
TRIGger:ACQuire[:IMMediate] (@<chanlist>)
This command sends an immediate trigger to the measurement
system. When the trigger system is initiated, a measurement trigger
causes the power system to measure the output voltage or current
and store the results in a buffer. The measured quantity, voltage or
current is specified by the SENSe:FUNCtion command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals