6 Language Dictionary
108 Series N6700 User’s Guide
TRIGger:ACQuire:SOURce BUS|PIN<pin>|TRANsient<chan>, (@<chanlist>)
TRIGger:ACQuire:SOURce? (@<chanlist>)
This command selects the trigger source for the measurement trigger
system. The following trigger sources can be selected:
BUS
GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
PIN<pin> Selects an output port connector pin. Pins 1 – 3 can be configured as external
trigger sources. [SOURce:]DIGital:PIN<n>:FUNCtion programs the function of
each pin. [SOURce:]DIGital:PIN<n>:POLarity programs the polarity of each pin.
TRANsient<chan> Selects the transient system of an output channel as the trigger source. To
generate triggers from the transient system use: [SOURce:]STEP:TOUTput,
[SOURce:]LIST:TOUTput:BOSTep, and [SOURce:]LIST:TOUTput:EOSTep.
The *RST value = BUS.
TRIGger:TRANsient[:IMMediate] (@<chanlist>)
This command generates an immediate transient trigger regardless of
the selected trigger source. Output triggers affect the following
functions: voltage, current, and current limit. You must initiate the
output trigger system before you can send any triggers.
When sent, the output trigger will:
Initiate an output change as specified by the Current Triggered or
Voltage Triggered commands.
Clears the WTG-tran bit in the Status Operation Condition
register after the transient trigger sequence has completed.
TRIGger:TRANsient:SOURce BUS|PIN<pin>|TRANsient<chan>, (@<chanlist>)
TRIGger:TRANsient:SOURce? (@<chanlist>)
This command selects the trigger source for the output trigger
system. The following trigger sources can be selected:
BUS GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
PIN<pin> Selects an output port connector pin. Pins 1 – 3 can be configured as external
trigger sources. [SOURce:]DIGital:PIN<n>:FUNCtion programs the function of
each pin. [SOURce:]DIGital:PIN<n>:POLarity programs the polarity of each pin.
TRANsient<chan> Selects the transient system of an output channels as the trigger source. To
generate triggers from the transient system use: [SOURce:]STEP:TOUTput,
[SOURce:]LIST:TOUTput:BOSTep, and [SOURce:]LIST:TOUTput:EOSTep.
The *RST value = BUS.
*TRG
This command generates a trigger when the trigger subsystem has
BUS selected as its source. The command has the same affect as the
Group Execute Trigger (<GET>) command.