EasyManuals Logo

Agilent Technologies N6751A User Manual

Agilent Technologies N6751A
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
6 Language Dictionary
108 Series N6700 User’s Guide
TRIGger:ACQuire:SOURce BUS|PIN<pin>|TRANsient<chan>, (@<chanlist>)
TRIGger:ACQuire:SOURce? (@<chanlist>)
This command selects the trigger source for the measurement trigger
system. The following trigger sources can be selected:
BUS
GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
PIN<pin> Selects an output port connector pin. Pins 1 – 3 can be configured as external
trigger sources. [SOURce:]DIGital:PIN<n>:FUNCtion programs the function of
each pin. [SOURce:]DIGital:PIN<n>:POLarity programs the polarity of each pin.
TRANsient<chan> Selects the transient system of an output channel as the trigger source. To
generate triggers from the transient system use: [SOURce:]STEP:TOUTput,
[SOURce:]LIST:TOUTput:BOSTep, and [SOURce:]LIST:TOUTput:EOSTep.
The *RST value = BUS.
TRIGger:TRANsient[:IMMediate] (@<chanlist>)
This command generates an immediate transient trigger regardless of
the selected trigger source. Output triggers affect the following
functions: voltage, current, and current limit. You must initiate the
output trigger system before you can send any triggers.
When sent, the output trigger will:
Initiate an output change as specified by the Current Triggered or
Voltage Triggered commands.
Clears the WTG-tran bit in the Status Operation Condition
register after the transient trigger sequence has completed.
TRIGger:TRANsient:SOURce BUS|PIN<pin>|TRANsient<chan>, (@<chanlist>)
TRIGger:TRANsient:SOURce? (@<chanlist>)
This command selects the trigger source for the output trigger
system. The following trigger sources can be selected:
BUS GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
PIN<pin> Selects an output port connector pin. Pins 1 – 3 can be configured as external
trigger sources. [SOURce:]DIGital:PIN<n>:FUNCtion programs the function of
each pin. [SOURce:]DIGital:PIN<n>:POLarity programs the polarity of each pin.
TRANsient<chan> Selects the transient system of an output channels as the trigger source. To
generate triggers from the transient system use: [SOURce:]STEP:TOUTput,
[SOURce:]LIST:TOUTput:BOSTep, and [SOURce:]LIST:TOUTput:EOSTep.
The *RST value = BUS.
*TRG
This command generates a trigger when the trigger subsystem has
BUS selected as its source. The command has the same affect as the
Group Execute Trigger (<GET>) command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals