Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Use of the serial link
MODIFYING THE HEADER
CHARACTER
ORDER
PARAMETER
RESPONSE ACTION
HDR
<val>
val = decimal
value of the
ASCII code of
the header
character
between 0
and 127.
OK
or
ERR3
l Change the header character
in front of a response address.
l The factory configuration is the
decimal code 035 of the cha-
racter #.
l Memorization of the customer
configuration by the monitoring.
l The code 000 corresponds to
no header character.
Example :
The value 047, the decimal ASCII code of the character /.
command #002HDR047<CR>
response /002OK
command 002IDN<CR>
response /002M4ADP100X V1.XX
INTERPRETING RESPONSES
SUPPLIED BY THE SERIAL LINK
OK or a specific response to the order if everything is correct.
ERR0 : setting fault
ERR1 : order fault
ERR2 : parameter fault
ERR3 : context fault
ERR4 : checksum fault
NOTE: The remote control mode has priority on RS 232 or RS 485
commands.