Appendix B | B-1
5100 HD CONFIGURED FOR REFINERY FLARE GAS
The 5100 HD TDLAS analyzer may be ordered in a dual-sample-cell, dual-
range-H
2
S conguration as AMETEK part number 9000-161-VE. This version of
the analyzer includes two optical sampling cells:
Standard, two-pass sample cell for a measuring range in terms of percent
by volume (e.g. 0-10% or 0-60% vol/vol). The accuracy for this measure-
ment is +/- 2% of the range.
Multi-pass, “Herriott” sample cell for a measuring ppm ranges (e.g.
0-300 ppmV). The accuracy here is +/- 15 ppmV.
Two laser diodes are used – one for each measuring range. The 5100 HD em-
ploys the same laser device for both ranges, but dierent wavelengths are used
to ensure an optimal match of absorption strength and sample cell path length for
each range.
The following two pages show a general arrangement drawing for this congura-
tion of the 5100 HD. A heated sample line entry boot is used for both the process
gas inlet and return to process lines. Unlike other 5100 HD versions, there is no
membrane separator/lter included in the analyzer oven. Accordingly, there is no
drain port or valve. The only operational maintenance is cleaning of the sample
cells as necessary.