User Interface | 4-9
Test Cong
The Test Cong key is used to dene device and communication settings. See
below for information on each section.
Modbus
Selects the communication setup. The Modbus protocol transmission mode
implemented is Remote Terminal Unit (RTU) with the analyzer operating as a
slave device.
MODBUS
Time Out (ms)*
Baud Rate
Slave Address*
9600
19200
Menu for Modbus key.
Slave Address
Enter the Slave Address.
Baud Rate
Select the baud rate at which you transfer data (9600 or 19200). The default
is 9600.
Time Out (ms)
Enter the time out in milliseconds.
Modbus interface parameters must be set up to establish communication. Use the
Modbus registers to obtain specic input/output data.
Modbus Registers can be found at the end of this chapter.