4-8 | Model 5100 HD Analyzer
Analog Range
The Analog Range key denes your 4 to 20 mA output setting. Use this setting
to scale the 4 to 20 mA output. The 4 to 20 mA output is proportional to the read-
ings.
ANALOG
RANGE
MAX CONCENTRATION*
MIN CONCENTRATION*
Main menu for Analog Range key.
Max Concentration
Enter the reading equivalent to 20 mA.
Min Concentration
Enter the reading equivalent to 4 mA.
Example:
To scale the analog output to represent a 0 to 100 PPMV range enter the follow-
ing settings:
Max Concentration 100
Min Concentration 0