Installation & Start-Up | 3-21
Initial Start-Up
1. Open the main sample shuto valve at the source.
2. AMETEK recommends you adjust the sample inlet pressure to between -4.5
PSIG and 10 PSIG.
Maximum inlet pressure is 10 PSIG. Damage can occur if higher
pressures are introduced into the analyzer.
3. Turn on the power at the source. Boot-up information will appear on the
analyzer display followed by the operation screen. Lines 3 and 4 of the
display will ash various messages and alarm conditions at start-up.
4. When the sample ow starts, the analyzer will adjust to normal operating
mode.
5. Adjust the inlet valve so that approximately 2000 ml/min of natural gas is
owing through the analyzer.
6. Set the ow rate of the owmeter to about 1200 ml/min for air to correspond
to the natural gas ow rate of about 2000 ml/min.
7. Use the software to set up your analyzer parameters (See Chapter 4 of this
manual).